ValProbe RT Loggers with bendable sensors provide a vast temperature range from -85°C to + 400°C. They are available as single and dual bendable sensor.

RTD technology delivers unrivalled measurement accuracy and the new logger electronic design guarantees long battery life. Operator programmable sample rates starting at 1 second, allow collection of up to 100 000 data points per sensor.

The antenna diversity concept secures an unmatched RF communication reliability.

Features
  • Temperature range for logger sensor: -85°C to 400°C;

  • Single and dual bendable sensor available;

  • Sensor length 12, 24, 36”;

  • Sensor diameter 2.4mm; 0.095”.

Applications:

  • Steam Sterilizer;

  • Dry Heat Sterilizer;

  • Steam in Place (SIP);

  • Water Cascade/Fall Sterilizer;

  • Incubators;

  • Stability Chambers;

  • Freezers;

  • FreezeDryer/Lyophilisation;

  • Vessels.

Main characteristics
Sensor type

Single and dual sensor

Accuracy

- 85°C to 140°C, ±0.1°C; 140°C to 400°C ±0.25°C

Calibration

NVLAP/DAkkS calibration

Real-time clock accuracy

< 15 sec/day

Environmental temperature

- 85 °C to 140 °C

Sensing element

Precision platinum RTD

Verification

Automated user verification capability

Data storage

100.000 samples retained in non-volatile memory

Battery

Field replaceable - 3.6 V lithium

Tip diameter

0.095"; 2.4 mm

Logger dimensions

Hight: 1,9

Meassurement range

- 85 °C to 400 °C

Sensor lenght

12, 24, 36" inches; 305, 610, 915 mm

Sampling rate

Starting at 1 sec

Environmental pressure

0-5 bar absolute

Logger material

Stainless steel 316 L and peek

Environmental humidity

0-100 % condensing

Regulatory compliance

UL and CE