ValProbe RT Loggers provide a vast temperature range from -85°C to + 140°C.
RTD technology delivers unrivalled measurement accuracy and the new logger electronic design guarantees long battery life. Operator programmable sample rates starting at 1 second, allow collection of up to 100 000 data points per sensor.
The antenna diversity concept secures an unmatched RF communication reliability.
Temperature range for complete logger: -85°C to 140°C;
Single sensor only;
Sensor length 1.5, 3, 6, 9”;
Sensor diameter 3mm; 0.118”.
Applications:
Steam Sterilizer;
Dry Heat Sterilizers;
Steam in Place (SIP);
Water Cascade/Fall Sterilizer;
Incubators;
Stability Chambers;
Freezers;
FreezeDryer/Lyophilisation;
Vessels.
Main characteristics | |
Sensor type | Single sensor |
Accuracy | -85°C to 140°C, ±0.1°C |
Calibration | NVLAP/DAkkS calibration |
Real-time clock accuracy | < 15 sec/day |
Environmental temperature | -85°C to 140°C |
Sensing element | Precision platinum RTD |
Verification | Automated user verification capability |
Data storage | 100.000 samples retained in non-volatile memory |
Tip diameter | 0.118"; 3 mm |
Logger dimensions | Hight: 1,9 |
Meassurement range | -85°C to 140°C |
Sensor lenght | 1.5, 3, 6, 9" inches; 38, 76, 152, 229 mm |
Battery | Field replaceable - 3.6 V lithium |
Sampling rate | Starting at 1 sec |
Environmental pressure | 0-5bar absolute |
Logger material | Stainless steel 316 L and peek |
Environmental humidity | 0-100 % condensing |
Regulatory compliance | UL and CE |