Ultra low temperature data logging system ULT90-15

Photo: Ultra low temperature data logging system ULT90-15
Photo: Ultra low temperature data logging system ULT90-15

The system provides a complete stand-alone temperature mapping and validation solution for freezers to ensure uniform storage and testing conditions.

MadgeTech's Ultra Low Temperature Freezer Validation System can measure and log data as low as -86 °C. It can be placed in small and large freezers of varying configurations. The system provides a complete stand-alone temperature mapping and validation solution for freezers to ensure uniform storage and testing conditions. The ULT90 includes a USB docking station with 1,83 meters of cable (IFC300), 15 Cryo-Temp data loggers, and MadgeTech's Secure Software and IQ/OQ/PQ validation protocols, aiding compliance with FDA 21 CFR Part 11.

For larger freezers, loggers can be plotted in one graph to show variances in different parts of the freezer. Finally, as failure to monitor external freezer temperatures may invalidate internal freezer temperatures, an additional Cryo-Temp can be included for placement outside of the freezer.

TEMPERATURE
Temperature sensor

Thermistor

Measurement range

-86 °C ... +35 °C

Resolution

0.1 °C

Accuracy

±1.0 °C

GENERAL CHARACTERISTICS
Operating environment

-86 °C ... +35 °C; 0 %RH ... 100 %RH

Memory

32,767

Reading rate

1 reading every 5 second up to 1 reading every 30 minutes

Material

ABS plastic

Battery life

2 years typical at 25 °C, 1 min. reading intervals / 1 year typical at -80 °C, 1 min. reading intervals

Dimensions

83 x 54 x 11 mm

Required interface package

IFC300