ValProbe Cryo Logger

New
Photo: ValProbe Cryo Logger
Photo: ValProbe Cryo Logger

Temperature Logger with thin and very flexible sensor.

The new Cryo Logger with the very thin and flexible tip is the perfect datalogger for an extended temperature range from -85°C to +140°C. The Logger provides a single solution for a variety of applications, such as cryogenic chambers, Lyophilizers, ultralow freezers and other deep temperature applications.It is also perfect to be used in autoclaves and other high temperature applications

It is a single solution that provides significant cost savings and productivity gains. RTD Technology delivers unrivalled measurement accuracy. The new logger design improves the battery life by three times over the existing design.

The new flexible Cryo Logger is designed for easy on-site user verification between studies with the traditional IRTD and low temp CTR-80 Bath or LTR-90 dry well.

The small and flexible sensor of the logger, with a stainless steel tip, allows a usage in most of the Pharma applications and it can also be used in combination with very small vials.

ValProbe satisfies FDA Regulation 21 CFR Part 11 requirements for electronic signatures and records and complies with EN 554 for saturated steam sterilization.

Features:
  • Temperature range for complete Logger: -85°C to 140°C;

  • Battery Life: 3x better than standard ValProbe Logger;

  • 1 m long and 1.8 mm thin robust flexible sensor with 2.2 mm stainless steel tip.

Applications:

  • Cryogenic Vessels;

  • Freeze Dryer;

  • Freezers;

  • Incubators;

  • Warehouses;

  • Sterilizers/Autoclaves.

Main characteristics
Sensor type

Flexible teflon cable

Calibration

Factory calibrated (NVLAP/DAkkS accredited) with user calibration capability

Real-time clock accuracy

15 seconds per 24 hours (0.0174 %)

Environmental temperature

- 85 °C to 140 °C

Sensing element

Precision platinum RTD

Data storage

10,000 samples retained in non-volatile EEPROM memory

Logger dimensions

1 13/16 x 1 3/8 in diameter (46 mm x 35 mm)

Meassurement range

0°C to 140°C, ± 0.1°C

Battery

Field-replaceable 3.6 V lithium thionyl chloride

Sampling rate

1 second to 12 hours

Logger material

316 L stainless steel

Total length (sensor + cable)

40" (100 cm)

Cable diameter

0.07" (1.8 mm)

Part number

XCFVP-40

Stainless steel tip

length 1" (25 mm); diameter 0.087" (2.2 mm)