The HiTemp140-M12 is a rugged and versatile high temperature data logger featuring an M12 probe connector. Compatible with a wide variety of M12 RTD probe styles, this data logger is capable of measuring up to +850 °C.
The datalogger is supplied with a factory calibration certificate ISO 17025.
The HiTemp140-M12 is a rugged and versatile high temperature data logger featuring an M12 probe connector. Compatible with a wide variety of M12 RTD probe styles, this data logger is capable of measuring up to +850 °C (probe dependent).
All of MadgeTech’s HiTemp140 data loggers have a water tight body enclosure made of food grade stainless steel and can be placed in environments with temperatures as high as +140 °C. The operating range of the data logger body can be extended to higher temperatures when used with MadgeTech’s thermal barrier options and suitable probe.
The HiTemp140-M12 records and stores up to 43,690 time stamped readings and is equipped with non-volatile solid state memory which retains data even if the battery becomes discharged.
A wide selection of compatible M12 probes are available on the market to choose from, making the HiTemp140-M12 the most dynamic data logger of its kind. With the ability to change probes as needed, this device satisfies a multitude of application needs with one powerful data logger versus the need for multiple loggers.
TEMPERATURE | |
Temperature sensor | Various RTD probes with M12 connector (sold separately) |
Measurement range | 18 Ω … 400 Ω; -200 °C … +850 °C (probe dependent) |
Resolution | 0.0001 Ω, 0.01 °C (probe dependent) |
Accuracy | ±0.015 Ω, ±0.05 °C (probe dependent). Calibrated accuracy range: *18 Ω ... 200 Ω; -200 °C ... +265 °C. Calibrated accuracies based on standard MadgeTech calibrations 0 Ω ... 200 Ω |
GENERAL CHARACTERISTICS | |
Operating environment | -40 °C … +140 °C; 0 %RH … 100 %RH non-condensing; 0.002 PSIA … 60 PSIA |
Memory | 43,690 |
Reading rate | 4 readings per second up to 1 reading every 24 hours |
Material | 316 Stainless Steel |
IP rating | IP40 - logger alone. The logger will inherit IP rating of the attached probe (up to, and including IP68/IP69K). |
Submersible | Yes |
Battery life | 2 years typical (1 minute reading rate) User Replaceable Battery |
Dimensions | 54.5 x 24.6 x 24.6 mm |
Required interface package | IFC400 / IFC406 |