Thermal shield for the HiTemp140 data logger.
Thermal shield for the HiTemp140 data logger. Available with a vented top or flush top. The HiTemp140-TS is a thermal shield designed for use with the HiTemp140 data logger. This durable enclosure allows the data logger to be used in temperatures from –200 °C to +250 °C, making it ideal for peanut roasting, food processing, autoclave validation and other extreme temperature applications.
Maximum exposure time chart |
||||
Ambient temperature |
HiTemp140-TS (Flush) |
HiTemp140-TS (Vented) |
||
Exposure time in air |
Exposure time in liquid |
Exposure time in air |
Exposure time in liquid |
|
-200 °C |
12 minutes |
N/A |
14 minutes |
N/A |
-180 °C |
13 minutes |
N/A |
15 minutes |
N/A |
-160 °C |
15 minutes |
N/A |
16 minutes |
N/A |
-140 °C |
17 minutes |
N/A |
18 minutes |
N/A |
-120 °C |
19 minutes |
N/A |
21 minutes |
N/A |
-100 °C |
22 minutes |
N/A |
24 minutes |
N/A |
-80 °C |
27 minutes |
N/A |
30 minutes |
N/A |
-60 °C |
37 minutes |
22 minutes |
42 minutes |
25 minutes |
-40 °C … +140°C |
Indefinitely |
Indefinitely |
Indefinitely |
Indefinitely |
+150 °C |
59 minutes |
34 minutes |
66 minutes |
40 minutes |
+160 °C |
51 minutes |
29 minutes |
57 minutes |
34 minutes |
+170 °C |
43 minutes |
25 minutes |
48 minutes |
29 minutes |
+180 °C |
37 minutes |
23 minutes |
42 minutes |
26 minutes |
+190 °C |
34 minutes |
20 minutes |
38 minutes |
23 minutes |
+200 °C |
31 minutes |
18 minutes |
34 minutes |
21 minutes |
+210 °C |
29 minutes |
17 minutes |
32 minutes |
19 minutes |
+220 °C |
27 minutes |
16 minutes |
30 minutes |
18 minutes |
+230 °C |
25 minutes |
15 minutes |
27 minutes |
17 minutes |
+240 °C |
23 minutes |
14 minutes |
26 minutes |
16 minutes |
+250 °C |
22 minutes |
13 minutes |
24 minutes |
15 minutes |