Thermal shield HiTemp140-TS

Photo: Thermal shield HiTemp140-TS
Photo: Thermal shield HiTemp140-TS
Photo: Thermal shield HiTemp140-TS
Photo: Thermal shield HiTemp140-TS

Thermal shield for the HiTemp140 data logger.

Thermal shield for the HiTemp140 data logger. Available with a vented top or flush top. The HiTemp140-TS is a thermal shield designed for use with the HiTemp140 data logger. This durable enclosure allows the data logger to be used in temperatures from –200 °C to +250 °C, making it ideal for peanut roasting, food processing, autoclave validation and other extreme temperature applications.

Maximum exposure time chart

Ambient temperature

HiTemp140-TS (Flush)

HiTemp140-TS (Vented)

Exposure time in air

Exposure time in liquid

Exposure time in air

Exposure time in liquid

-200 °C

12 minutes

N/A

14 minutes

N/A

-180 °C

13 minutes

N/A

15 minutes

N/A

-160 °C

15 minutes

N/A

16 minutes

N/A

-140 °C

17 minutes

N/A

18 minutes

N/A

-120 °C

19 minutes

N/A

21 minutes

N/A

-100 °C

22 minutes

N/A

24 minutes

N/A

-80 °C

27 minutes

N/A

30 minutes

N/A

-60 °C

37 minutes

22 minutes

42 minutes

25 minutes

-40 °C … +140°C

Indefinitely

Indefinitely

Indefinitely

Indefinitely

+150 °C

59 minutes

34 minutes

66 minutes

40 minutes

+160 °C

51 minutes

29 minutes

57 minutes

34 minutes

+170 °C

43 minutes

25 minutes

48 minutes

29 minutes

+180 °C

37 minutes

23 minutes

42 minutes

26 minutes

+190 °C

34 minutes

20 minutes

38 minutes

23 minutes

+200 °C

31 minutes

18 minutes

34 minutes

21 minutes

+210 °C

29 minutes

17 minutes

32 minutes

19 minutes

+220 °C

27 minutes

16 minutes

30 minutes

18 minutes

+230 °C

25 minutes

15 minutes

27 minutes

17 minutes

+240 °C

23 minutes

14 minutes

26 minutes

16 minutes

+250 °C

22 minutes

13 minutes

24 minutes

15 minutes